Start-Up and Use
For example, in Figure 11, the histogram is for the last 12 hours.
During those 12 hours, 68% of the %SpO
2 measurements were
from 96 to 100, 7% of the measurements were from 91 to 95, and
25% of the measurements were from 0 to 80.
Zoom
Pressing the ZOOM softkey changes the period of time indicated
on a trend graph. Selectable times displayed graphically are 24,
12, 8, 4, 2, or 1 hours, and 30 or 12 minutes. The location of the
cursor, as indicated by the time in the upper right-hand corner of
the screen, remains the same.
A table is used to display 20-second trends.
The ZOOM softkey is not displayed for the histogram trend view.
20-second trends table
The 20-second trends table is available by continuing to press the
ZOOM softkey. As indicated in Figure 12, the time indicated on
trend graphs by the cursor is displayed in the upper right-hand
corner of the screen and is highlighted in the table. Press the
Down Arrow button to highlight the time of the trends readings.
The readings are in 2-second increments and go back in time as
the Down Arrow button is pressed and forward in time as the Up
Arrow button is pressed.
20SEC TREND
%SPO2
TIME
TIME
BPM
28JUL98
21:31:38
21:31:36
21:31:34
21:31:32
21:31:30
96
97
97
97
97
78
79
78
78
78
%SPO2 BPM
21:31:46
21:31:44
21:31:42
21:31:40
97
97
97
96
96
78
79
80
78
78
21:31:48
21:31:48
Figure 12: 20-Second Trends
When the oldest reading on the screen is highlighted and the
Down Arrow button is pressed again, the screen shifts to display
an older column of readings. Similarly, if the newest reading on
the screen is highlighted and the Up Arrow button is pressed, the
screen shifts to display a newer column of readings.